1.4: Spectral Techniques
- Page ID
- 38584
\( \newcommand{\vecs}[1]{\overset { \scriptstyle \rightharpoonup} {\mathbf{#1}} } \)
\( \newcommand{\vecd}[1]{\overset{-\!-\!\rightharpoonup}{\vphantom{a}\smash {#1}}} \)
\( \newcommand{\id}{\mathrm{id}}\) \( \newcommand{\Span}{\mathrm{span}}\)
( \newcommand{\kernel}{\mathrm{null}\,}\) \( \newcommand{\range}{\mathrm{range}\,}\)
\( \newcommand{\RealPart}{\mathrm{Re}}\) \( \newcommand{\ImaginaryPart}{\mathrm{Im}}\)
\( \newcommand{\Argument}{\mathrm{Arg}}\) \( \newcommand{\norm}[1]{\| #1 \|}\)
\( \newcommand{\inner}[2]{\langle #1, #2 \rangle}\)
\( \newcommand{\Span}{\mathrm{span}}\)
\( \newcommand{\id}{\mathrm{id}}\)
\( \newcommand{\Span}{\mathrm{span}}\)
\( \newcommand{\kernel}{\mathrm{null}\,}\)
\( \newcommand{\range}{\mathrm{range}\,}\)
\( \newcommand{\RealPart}{\mathrm{Re}}\)
\( \newcommand{\ImaginaryPart}{\mathrm{Im}}\)
\( \newcommand{\Argument}{\mathrm{Arg}}\)
\( \newcommand{\norm}[1]{\| #1 \|}\)
\( \newcommand{\inner}[2]{\langle #1, #2 \rangle}\)
\( \newcommand{\Span}{\mathrm{span}}\) \( \newcommand{\AA}{\unicode[.8,0]{x212B}}\)
\( \newcommand{\vectorA}[1]{\vec{#1}} % arrow\)
\( \newcommand{\vectorAt}[1]{\vec{\text{#1}}} % arrow\)
\( \newcommand{\vectorB}[1]{\overset { \scriptstyle \rightharpoonup} {\mathbf{#1}} } \)
\( \newcommand{\vectorC}[1]{\textbf{#1}} \)
\( \newcommand{\vectorD}[1]{\overrightarrow{#1}} \)
\( \newcommand{\vectorDt}[1]{\overrightarrow{\text{#1}}} \)
\( \newcommand{\vectE}[1]{\overset{-\!-\!\rightharpoonup}{\vphantom{a}\smash{\mathbf {#1}}}} \)
\( \newcommand{\vecs}[1]{\overset { \scriptstyle \rightharpoonup} {\mathbf{#1}} } \)
\( \newcommand{\vecd}[1]{\overset{-\!-\!\rightharpoonup}{\vphantom{a}\smash {#1}}} \)
\(\newcommand{\avec}{\mathbf a}\) \(\newcommand{\bvec}{\mathbf b}\) \(\newcommand{\cvec}{\mathbf c}\) \(\newcommand{\dvec}{\mathbf d}\) \(\newcommand{\dtil}{\widetilde{\mathbf d}}\) \(\newcommand{\evec}{\mathbf e}\) \(\newcommand{\fvec}{\mathbf f}\) \(\newcommand{\nvec}{\mathbf n}\) \(\newcommand{\pvec}{\mathbf p}\) \(\newcommand{\qvec}{\mathbf q}\) \(\newcommand{\svec}{\mathbf s}\) \(\newcommand{\tvec}{\mathbf t}\) \(\newcommand{\uvec}{\mathbf u}\) \(\newcommand{\vvec}{\mathbf v}\) \(\newcommand{\wvec}{\mathbf w}\) \(\newcommand{\xvec}{\mathbf x}\) \(\newcommand{\yvec}{\mathbf y}\) \(\newcommand{\zvec}{\mathbf z}\) \(\newcommand{\rvec}{\mathbf r}\) \(\newcommand{\mvec}{\mathbf m}\) \(\newcommand{\zerovec}{\mathbf 0}\) \(\newcommand{\onevec}{\mathbf 1}\) \(\newcommand{\real}{\mathbb R}\) \(\newcommand{\twovec}[2]{\left[\begin{array}{r}#1 \\ #2 \end{array}\right]}\) \(\newcommand{\ctwovec}[2]{\left[\begin{array}{c}#1 \\ #2 \end{array}\right]}\) \(\newcommand{\threevec}[3]{\left[\begin{array}{r}#1 \\ #2 \\ #3 \end{array}\right]}\) \(\newcommand{\cthreevec}[3]{\left[\begin{array}{c}#1 \\ #2 \\ #3 \end{array}\right]}\) \(\newcommand{\fourvec}[4]{\left[\begin{array}{r}#1 \\ #2 \\ #3 \\ #4 \end{array}\right]}\) \(\newcommand{\cfourvec}[4]{\left[\begin{array}{c}#1 \\ #2 \\ #3 \\ #4 \end{array}\right]}\) \(\newcommand{\fivevec}[5]{\left[\begin{array}{r}#1 \\ #2 \\ #3 \\ #4 \\ #5 \\ \end{array}\right]}\) \(\newcommand{\cfivevec}[5]{\left[\begin{array}{c}#1 \\ #2 \\ #3 \\ #4 \\ #5 \\ \end{array}\right]}\) \(\newcommand{\mattwo}[4]{\left[\begin{array}{rr}#1 \amp #2 \\ #3 \amp #4 \\ \end{array}\right]}\) \(\newcommand{\laspan}[1]{\text{Span}\{#1\}}\) \(\newcommand{\bcal}{\cal B}\) \(\newcommand{\ccal}{\cal C}\) \(\newcommand{\scal}{\cal S}\) \(\newcommand{\wcal}{\cal W}\) \(\newcommand{\ecal}{\cal E}\) \(\newcommand{\coords}[2]{\left\{#1\right\}_{#2}}\) \(\newcommand{\gray}[1]{\color{gray}{#1}}\) \(\newcommand{\lgray}[1]{\color{lightgray}{#1}}\) \(\newcommand{\rank}{\operatorname{rank}}\) \(\newcommand{\row}{\text{Row}}\) \(\newcommand{\col}{\text{Col}}\) \(\renewcommand{\row}{\text{Row}}\) \(\newcommand{\nul}{\text{Nul}}\) \(\newcommand{\var}{\text{Var}}\) \(\newcommand{\corr}{\text{corr}}\) \(\newcommand{\len}[1]{\left|#1\right|}\) \(\newcommand{\bbar}{\overline{\bvec}}\) \(\newcommand{\bhat}{\widehat{\bvec}}\) \(\newcommand{\bperp}{\bvec^\perp}\) \(\newcommand{\xhat}{\widehat{\xvec}}\) \(\newcommand{\vhat}{\widehat{\vvec}}\) \(\newcommand{\uhat}{\widehat{\uvec}}\) \(\newcommand{\what}{\widehat{\wvec}}\) \(\newcommand{\Sighat}{\widehat{\Sigma}}\) \(\newcommand{\lt}{<}\) \(\newcommand{\gt}{>}\) \(\newcommand{\amp}{&}\) \(\definecolor{fillinmathshade}{gray}{0.9}\)- If one switched from a large cuvette to a smaller one, the %T readings on the spectrophotometer would:
- decrease
- increase
- fluctuate
- remain unchanged
- have equal absorbance readings
- A solution is measured in a spectrophotometer with a resulting %T of 10%. Calculate the absorbance.
- 0
- 0.5
- 1.0
- 1.5
- 2.0
- Compound Y has been determined to have a molar absorptivity (extinction coefficient) of 1000 liters/ mole • cm at 390 nm. An unknown solution of this compound has an absorption of 0.750 when placed in a cell with a light path of 1 cm. The concentration of this compound is:
- 375 mole/L
- 750 mole/L
- 750 \(\mu\)mole/L
- 0.375 mmole/L
- 750 mmole/L
- The %T scale on a spectrophotometer is a logarithmic one while the absorbance scale is a linear scale. Because of this relationship, very high absorbance readings are inaccurate.
- true
- false
- A standard solution of Compound X at 10 mg/L gives A450 = 1.200 when used in a specific colorimetric assay. A serum sample analyzed for Compound X gives the following results: run undiluted: A450 = 2.400, run diluted 1:10: A450 = 0.300. The concentration of Compound X in serum is:
- 2.5 mg/L
- 2.0 mg/L
- 20 mg/L
- 25 mg/L
- 200 mg/L
- Compounds A and B both have a chemical formula of C9H18N2O; therefore, both compounds must have the same molar absorptivity coefficient (\(\epsilon\) or a) and the same spectral curve:
a. true
b. false - Refer to the following spectral-transmission curve made on the same colored solution using two different instruments. All of the following statements are true except: (Note:one instrument had a bandpass of 35 nm, the other 5 nm)
- A is the curve which was obtained with the 35 nm instrument
- B is the curve obtained with the 5 nm instrument
- reading the color intensity at point C would yield the greatest sensitivity
- 550 nm would be an appropriate wavelength at which to read the reaction
- a and b
- To detect transmitted light, either barrier tubes or photomultiplier tubes can be
used. Both work on the principle that incident photons of light interact with a thin
metallic layer resulting in the generation of an electric current.- true
- false
Instruments - Spectrophotometers
Match each of the following instrument components with its most accurately described function — Questions: 53 thru 55
- slit(s)
- barrier layer cell
- interference filter
- photomultiplier
- potentiometer
- Converts electromagnetic energy to electricity
- Renders light rays parallel
- Converts light to electricity and amplifies the electron flow
- Which of the following combinations are not consistent?:
- tungsten lamp, diffraction grating, glass cuvette, barrier layer cell
- Xenon arc lamp, diffraction grating, quartz cuvette, photomultiplier tube
- Deuterium lamp, quartz prism, quartz cuvette, photomultiplier tube
- Deuterium lamp, glass prism, glass cuvette, photomultiplier tube
- B and D
- The principle of atomic absorption spectrophotometry is best stated as:
- ground state (unexcited) atoms absorb light energy of a characteristic wavelength
- excited atoms emit light of a characteristic wavelength after absorbing energy
- excited atoms emit light of a longer wavelength after absorption of energy of
a shorter wavelength - ground state (unexcited) atoms absorb light energy of a characteristic wavelength, become ionized, and emit light
- excited atoms absorb light
- The most common light source in atomic absorption instruments is a:
- Xenon arc
- flame
- hollow cathode lamp
- grating
- deuterium lamp
- In atomic absorption spectrophotometry, the purpose of the light source is:
- to provide a chamber to house the flame
- to produce a wavelength of light characteristic for the metal in the cathode
- to act as a detector of light energy transmitted by the flame
- to transmit argon or neon gas into the flame as a source of fuel
- to act as a detector of light energy allowed to be transmitted by the chopper
- If one desired to measure calcium by atomic absorption spectrophotometry, the light source would have to be made of which component(s)?:
- magnesium
- deuterium
- tungsten and quartz
- calcium
- zinc
- The heat of the atomic absorption (A.A.) flame is not critical for accurate and precise measurement of Ca++ since the procedure uses a calibration curve.
- True
- False
- A single source for the hollow-cathode lamp is used for Ca++, Mg++, and Li+, while a different lamp must be used to measure Cu++.
- True
- False
- Most atomic absorption instruments use filters for their monochromometers.
- True
- False
- The purpose of the flame in atomic absorption is to excite the atoms so that they can absorb light from the hollow cathode lamp.
- True
- False
- Nephelometers measure light at:
- a different wavelength than the input light
- the wavelength of maximum
- the wavelength of maximum fluorescence
- the same wavelength as the input light
- the point of longest scatter
- Turbidity measurement is equivalent to:
- fluorescence
- nephelometry
- absorbance
- phosphorescence
- quenching
- Most spectrophotometric analyses use absorbance measurements because there is no direct relationship between %T and concentration.
- True
- False
Match the principle which best describes the effect of each of the following (questions 24 thru 28):
Principle
- spectrophotometry
- fluorometry
- turbidity
- nephelometry
Effect
- Increase in light absorption caused by particulate interaction with absorbent compound.
- Absorption and emission of light at different wavelengths.
- Decrease in observed light intensity caused by light scattering.
- Increase in observed light intensity caused by light scattering.
- Absorption of incident light by compound resulting in a decrease in observed light intensity
- Which of these reactions can occur in a flame? (M = metal)
- M+ + electron → M°
- M* → M° + photon
- M° + photon → M*
- a and c
- all of the above
- A spectral transmittance curve for compound X is shown below. The wavelength at which measurement should be made is:
- 400
- 450
- 500
- 550
- 600
- When an atom, ion or molecule absorbs a photon of light, this is termed:
- atomic transition
- fluorescence
- radiationless transition
- electronic transition
- Beer’s Law
- A beam of white light is passed through a green filter, and the transmitted light is passed into a green solution. The transmitted light from this solution was measured and the %T and absorbance recorded. This experiment is repeated. The filter remains green but now a red solution is used. Which of the following responses for %T and absorbance would be expected?
- %T and absorbance both increase
- %T increases, absorbance decreases
- %T decreases, absorbance increases
- %T and absorbance both decrease
- %T and absorbance values remain unchanged
- Both atomic absorption and flame photometers use filters as the monochrometers:
- true
- false
- In sunlight, what color is the solution of the compound whose spectral curve is shown below:
- blue
- green
- yellow
- purple
- colorless
- 10 mL of a solution is read spectrophotometrically in a 2 cm cuvette. The
absorbance obtained after correcting for dilution is 30. The molar absorptivity of
this compound, whose M.W. is 1000, is 150. What is the concentration of this
compound in mg/mL?- 0.1 mg/mL
- 1 mg/mL
- 10 mg/mL
- 100 mg/mL
- 1000 mg/mL
- A 0.01 M solution of a compound is prepared and its absorption at 260 nm is
measured in a 20 mm cell and found to be 0.500. What is molar absorptivity of
this compound?- 50
- 25
- 12.5
- 2.5
- none of the above
- You are performing a manual procedure in which you need 14 10 mm cuvettes. Only 13 10 m cuvettes are available so you read one of the standards in a 20 mm cuvette. Upon calculating the concentration of this solution, you find the standard appears:
- more concentrated than it should be
- less concentrated than it should be
- the same concentration as expected
- A linearity check is performed on the spectrophotometer using a solution that follows Beer’s Law up to a concentration of 2.5. The following values are obtained:
Concentration (mol/L) |
Absorbance |
---|---|
1.01 | 1.60 |
0.75 | 1.35 |
0.50 | 0.90 |
0.25 | 0.45 |
You would conclude:
- the instrument is linear
- the instrument is non-linear
Use the following code to answer questions 39 thru 41:
- If A is greater than B
- B is greater than A
- A and B are about equal
- A. U.V. light absorbed by glass
B. U.V. light absorbed by fused silica - A. linearity of spectrum produced by a prism
B. linearity of spectrum produced by a diffraction grating - A. transmission of blue light by a cuvette filled with a red solution
B. transmission of yellow light by a cuvette filled with a red solution
Use the following Key to answer questions 42 through 56:
- 1,2 and 3 are correct
- 1 and 3 are correct
- 2 and 4 are correct
- only 4 is correct
- all are correct
- The following can affect the linear response of concentration vs. absorbancy and thus result in a deviation from Beer’s Law:
- stray light
- use of photometer instead of a spectrophotometer
- very high analyte concentration
- Using only one standard for calculation
- On linear graph paper, a straight line through zero with three plotted standards indicates which of the following?:
- the concentration of the standard vs. the absorbance is linear
- the test is in compliance with Beer’s Law’
- the concentration of the standard vs. the log %T is linear
- the test is not in compliance with Beer’s Law.
- Deviations from Beer’s Law are caused by
- stray light
- polychromatic light
- very high concentrations of substance being measured in a colorimetric reaction
- very low concentration of absorbing material
- Which of the following are preferred for checking the wavelength calibration of a spectrophotometer?:
- cobalt sulfate
- didymium jilter
- potassium dichromate
- helium oxide filter
- The concept of thermal excitation as used in emission photometry involves:
- the absorption of light by certain alkali metals
- characteristic line spectra being given off by excited orbital electrons as they return to the ground state
- orbital electrons absorbing radiant energy which is converted to electrical energy as measured by a gal manometer
- the heating of alkali metal salts to produce a colored light of characteristic wavelength
- In flame photometry the measured light comes from:
- burning fuel
- light produced when atoms go from ground state (M°) to excited state (M*)
- light from a tungsten bulb
- light produced when atoms go from excited state (M*) to ground state (M°)
- The purpose of the nebulizer in atomic absorption is to:
- break chemical bonds
- excite atoms
- measure the sample flow rate
- disperse aqueous sample into fine droplets
- Lanthanum diluents is used in atomic absorption measurements to:
- precipitate interfering proteins
- bind Mg++ so it will not interfere with Ca++ measurements
- release Ca++
- release Ca++ complexes
- Which of the following statement(s) concerning the use of cesium (Cs) as an internal standard for flame photometry is (are) true? Cesium:
- is normally present in very low concentrations in body fluids
- has an emission line easily distinguished from those of Na and K
- is used to correct for minor variations of reaction conditions
- can be replaced by other elements such as magnesium
- Turbidimetry and nephelometry measure:
- fluorescence in turbid solution
- light absorbance by particles in suspension
- light absorbance by a colored substance
- light scatter by particles in suspension
- Components of a nephelometer include:
- light source
- detector
- filter
- chopper
- In fluorometry:
- the compound measured absorbs light of a shorter wavelength and emits light of a longer wavelength
- quenching occurs when a substance other than the one being measured emits fluorescence
- fluorometery is a relative measurement of light signal
- the compound measured absorbs light of a longer wavelength and emits light of a shorter wavelength
- Which of the following types of lamps would be most appropriate for the entire UV range?
- mercury vapor
- tungsten iodide with quartz envelope
- tungsten with glass envelope
- deuterium
- An acceptable means of achieving spectral isolation in the UV range is (are):
- reflectance diffraction grating
- glass filter
- quartz prism
- glass prism
- The correct wavelength for reading a sample on a spectrophotometer is:
- the wavelength of the highest absorbance maxima
- the wavelength at which the most light passes through the solution
- the wavelength at which a straight line (percent transmittance vs. concentration) is obtained on a semi-log paper
- the wavelength at which the photo detector has optimum response.
- When a fluorescent molecule absorbs polarized excitation light, the emitted light loses polarization when:
- the the light intensity decreases
- light intensity increases
- the excited state changes
- the fluorescent molecule is held rigid
- the fluorescent molecule
- To retain a linear relationship between fluorescence intensity and the amount of light absorbed by a solution, the absorbance of the solution should not exceed:
- 0.001
- 0.01
- 0.1
- 1.0
- 2.0
- Fluorescence polarization assays are different from those that measure only fluorescence intensity because:
- those which measure intensity can tolerate unstable light sources.
- those which measure intensity can tolerate less intense light.
- the light intensity is greater.
- polarization measurements are less affected by variations in fluorescence intensity.
- polarization measurements are more affected by variations in fluorescence intensity.
- Florescence attenuation assays are used in dedicated fluorometric instruments such as the Abbott TDx. In these assays, the fluorescent dye concentration is:
- Constant
- linearly proportional to analyte Concentration
- logarithmically proportional to analyte concentration
- proportional to analyte formation
- inversely proportional to analyte formation
- Chemiluminescent reactions are those in which one of the products is
- Luminol
- oxidized luminol
- reduced luminol
- heat
- light
- Time delayed fluorescence is a term that refers to:
- fluorescence which starts to occur only after one millisecond following exposure to exciting light
- fluorescence which is delayed by solution chemistry
- fluorescence which is measured after a time delay
- the time it takes for a photon to jump from a ground state to an excited state
- the delayed time required before the incident light excites the fluor
- Atomic absorption can be considered the opposite of flame emission because in atomic absorption:
a. thermal energy applied to the sample causes formation of excited atoms
b. a specific light source is necessary for the determination of each element
c. electron’s movement to a higher energy level is measured.
d. the amount of light absorbed by an electron moving to a higher energy level is measured.
e. light is given off by excited atoms. - In a fluorometer, when the primary excitation source produces fluorescence, the secondary light emission by the sample has:
a. a longer wavelength than the primary light
b. a shorter wavelength than the primary light
c. the same wavelength as the primary light
d. no specific wavelength
e. any of the above - Which of the following describes a physical property that distinguishes luminescence from fluorescence?
a. fluorescence is at longer wavelengths.
b. in luminescence a photon goes from an excited state to a ground state.
c. in fluorescence a photon goes from an excited state to a ground state.
d. luminescence has a longer transition time to the ground state.
e. fluorescent light is always polarized. - Chemiluminescence can be induced by
- light
- flame ionization
- a chemical reaction
- firefly luciferase
- atomic absorption
- Answer
-
- b (p. 88)
- c (p. 88,38)
- c (p. 88,38)
- b (p. 88)
- d (p. 88,38)
- b (p. 86-88)
- c (p. 89,90,92)
- a (p. 91)
- b (p. 91)
- a (p. 91)
- d (p. 91)
- d (p. 91-93)
- a (p. 95)
- c (p. 95)
- b (p. 95)
- d (p. 95-96)
- b (p. 97)
- a (p. 96)
- b (p. 96)
- b (p. 95)
- d (p. 101-103)
- c (p. 101-102)
- a (p. 88)
- c (p. 101-102)
- b (p. 97-98)
- c (p. 101-102)
- d (p. 101-103)
- a (p. 88-89)
- e (p. 97)
- c (p. 92-93)
- d (p. 86)
- c (p. 86-88)
- b (p. 95-97)
- d (p. 86)
- d (p. 88-89)
- b (p. 88-89)
- a (p. 88-89)
- b (p. 88-89)
- 1 (p. 91)
- 2 (p. 90)
- 3 (p. 88)
- b (p. 88-89)
- a (p. 88-89)
- a (p. 88-89)
- c (p. 93-94)
- c (p. 87, 97)
- d (p. 97)
- d (p. 96)
- d (p. 96)
- a (p. 97)
- d (p. 101-102)
- a (p. 102)
- b (p. 97-98)
- d (p. 89-90)
- b (p. 90)
- b (p. 92-93)
- e (p. 100-101)
- c (p. 99)
- d (p. 100-101)
- a (p. 99)
- e (p. 99-100)
- c (p. 99)
- d (p. 99)
- a (p. 98)
- d (p. 98-100)
- c (p. 99-100)